Search
Close this search box.

CIC-D150 Ion Chromatograph

Leakage alarm

If there is liquid leakage in the pipeline, the D150 liquid leakage detector will detect the liquid, and a red prompt sign will appear on the computer and the touch screen, and an alarm sound will be given to remind in time, and the pump will stop automatically after 5 minutes of no treatment.

Auto-range

When D150 ion chromatograph is operated, it is easy to realize the simultaneous determination of 5ppb-100ppm concentration sample without setting the range, and the signal is displayed with digital signal μs/cm.

Gas liquid separator

The bubble in the eluent will increase the baseline noise and reduce the sensitivity. A micro gas-liquid separator is set in the pipeline between the infusion pump and the eluent bottle to separate the bubble in the eluent from the eluent.

 Timing startup preheating

It usually takes about 1 hour for the ion chromatograph to balance the system from start-up to sample injection analysis. When the user has prepared the eluent (or pure water for the eluent), he can set the start-up running time of the instrument in advance (the maximum setting is 24 hours), complete the start-up operation and all parameter settings.

 Intelligent maintenance

Set “intelligent maintenance”, the instrument can complete the flow path switch to the pure water path, the flow rate is set to 0.5ml/min, running for 1 hour.

 Mobile APP

Mobile app is easy to operate. APP monitoring: put the device in your pocket, no matter where you are, turn on your mobile phone to view and control the field device. The mobile app can remotely control the instrument on / off and observe the operation performance parameters of the instrument.

Intelligent large screen

The large screen displays the operation parameters and status of the instrument, which is convenient for the operator to check the equipment status on site, and to complete the operation of instrument on-off, instrument maintenance, etc

Parameters

Pump

Maximum Pressure:42MPa

Pressure Display Accuracy:≤0.1MPa

Flow Rate Range:0.001~9.999mL/min

Injector

Maximum Pressure:35MPa(PEEK)

Conductivity Detector

Cell Volume: ≤0.8μL

Detection Range:0~45000μS/cm 

Baseline Noise:≤0.001μS

Baseline Drift:≤0.001μS

Detection Resolution: ≤0.0020nS/cm              

Electronic Noise:0.02nS

Temperature Range:Room temperature+5~60℃(41 ~140℉)

Maximum Pressure:10.0MPa 

Power Requirements:90W 

Dimensions(L*W*H*):336*650*458(mm)

Net Weight(KGS):26

Gross Weight(KGS):32 

Request a Quote!