Non Aqueous solution (Ag/Ag+ type)

  • For apply in an organic solvent sample
  • Internal solution changeable

∗RE-7VT Non Aqueous reference electrode screw type (Ag/Ag+) was designed to be used in ALS brand flow cell (CrossRadialEQCMSEC-2F), for this reason there is a possibility that it will not be compatible in another brand flow cell.

Manufactured by: ALS, Japan

Description

  • For apply in an organic solvent sample
  • Internal solution changeable

Preservation of the electrode

  • Fill the sample holder with an acetonitrile/0.1 M tetrabutylammonium perchlorate/0.01 M AgNO3 solution.
  • Rinse the Ag wire with acetone.
  • Keep the electrode immersed in an acetonitrile/0.1 M tetrabutylammonium perchlorate solution, with or without Ag ion (0.01 M AgNO3).

Note:

  1. Avoid the dry of the Vycor® glass, it could cause the crash of the Vycor® glass tip, caused by the salt crystallization inside of the glass. Make sure to keep the tip always wet.
  2. Before assembly the electrode, immerse and keep the Vycor® glass tip in an organic solvent for at least one hour.
  3. If bubbles are present around the Vycor® glass, remove with a slight tapping. Air bubbles will isolate, blocking the current flow.
  4. Use only high grade products and reagents.
  5. Wash with acetonitrile. Salt crystallization, in the inner side of the Teflon cap and at the edge of the sample holder, makes difficult the assembly.

Line Up

Catalog No.DescriptionJunctionElectrolytePurpose
12171RE-7 Non Aqueous reference electrode (Ag/Ag+)IPPGACN/TBAPSVC-2, SVC-3, VC-4, SBC Bulk electrolysis, RRDE, QCM
13394RE-7S Non Aqueous reference electrode (Ag/Ag+)IPPGACN/TBAPSECM
13489RE-7VT Non Aqueous reference electrode screw type (Ag/Ag+)CeramicsACN/TBAPFlow cell (Cross, Radial, EQCM, SEC-2F)
Optional items (sold separately)
12549RE-7 Electrolyte solution (10 mL)ACN/TBAPFor refill of RE-7 line up (about 15 times)
12108RE-PV Preservative vial for reference electrodeFor preservation of OD 6.0 mm reference electrode
12057RE-7 Teflon cap with Ag wireWith Ag wire
12058RE-7S Teflon cap with Ag wireWith Ag wire
12176Sample holder dia 6 mm (2 pcs)IPPGFor double junction, manufacturing of the electrode
12177Sample holder dia 9 mm (2 pcs)IPPGFor double junction, manufacturing of the electrode
ACN: acetonitrile / TBAP: tetrabutylammonium perchlorate
IPPG: Ion Permeability Porous Glass

Precautions

Electrode

This electrode is silver-silver ion reference electrode for non-aqueous solution.
In order to reduce the liquid junction potential, use the same solvent and supporting electrolyte, as a sample and electrode internal solution.

Solvent

The solvent in non-aqueous electrochemical should be considered as following:

  1. Electrode activity material and produced substance solubility.
  2. Select according to the potential area to be measured.

For example, dimethyl sulfoxide (DMSO) is major solvent in the negative potential area, and acetonitrile (ACN) is major solvent in the positive potential area.
The non-aqueous solution could contains impurities, and the impurities could cause the block reactions. Distilled the solvent sufficiently, to eliminated impurities before using.

Supporting electrolyte

If the sample is dissolved in an organic solvent, the supporting electrolyte must be added.
In order to select the supporting electrolyte, it is necessary to consider the following:

  1. Solubility in organic solvents
  2. Wide potential window
  3. No reaction with organic solvent

Typical supporting electrolyte:

  • TEAP: Tetraethylammonium perchlorate
  • TBAPF6: Tetrabutylammonium hexafluorophosphate