The Model 4330 Voltammetric Analyzer is a combination of three electrochemical instruments in one.

 

Thanks to the evolution of electronics and firmware, Amel have been able to combine all modern voltammetric techniques with Potentiometric Stripping analysis (PSA) and pH measurement.


Trace analysis of heavy metals and organic compounds down to ppb (ug/L) levels will easily be
accomplished by the most appropriate technique  and electrode. The integrated cell stand offers
unrivaled versatility, thanks to the wide choice of batch/flow cells and electrodes, from classical
hanging mercury drop and solid electrodes to modern rotating disk and screen printed
electrodes.

Automatic quantitative analysis is then made easy by integration with an Amel 235 Automatic
Dispenser. This combination will allow you to perform routine analysis by a simple button touch
as well as research work on the same instrument.

Detection
AD Amperometric Detection
PD Potentiometric Detection
DSA Double Step Amperometry
DSV Double Step Potentiometry
PAD Pulsed Amperometric Detection
Voltammetric
LSV Linear Scan Voltammetry
CYV Cyclic Voltammetry
GLV Galvanostatic Linear Voltammetry
GCV Galvanostatic Cyclic Voltammetry
SWV Square Wave Voltammetry
NPV Normal Pulse Voltammetry
ACV Alternating Current Voltammetry
DPV Differential Pulse Voltammetry
DNV Differential Normal Pulse Voltammetry
DAV Differential Alternate Pulse Voltammetry
Stripping
LSS Linear Scan Stripping
ACS Alternate Current Stripping
SWS Square Wave Stripping
DAS Differential Stripping
DPS Differential Pulse Stripping
DNS Differential Normal Pulse Stripping
PSA Potentiometric Stripping Analysis
CCSA Constant Current Stripping Analysis

Specifications are subject to change without previous notice

Counter Electrode
Voltage Output ± 12V max
Current Output ± 12mA max
Slew Rate 1mV/s to 10V/s
Protection Thermal, overload and short-circuit
Working Electrode
Current Measure From 1nA to 10mA Full Scale in 8 ranges
Current Resolution From 1pA at 1nA Full Scale to 1μA at 10mA Full Scale
Measuring Accuracy ± 0,2% & 0,1% (conversion at Full Scale)
Reference Electrodes
Input Impedance > 1TΩ
Input Capacitance < 20pF (1m cable)
Biasing Current < 2pA at 25°C
Common Mode Rejection > 50dB full frequency response
Voltage Range ± 10V max
Polarization Capability
Voltage ± 4V max
Current ± 10mA max
Voltage Resolution 1mV
Current Resolution 1pA
Accuracy ± 0,2% & 0,1% (conversion at Full Scale)
Response Time
Potensiostatic Rise Time 10μs resistive load (1000Ω)
Galvanostatic Rise Time 25μs resistive load (1000Ω)
Meters and Interface
A/D Converter 16 BIT
D/A Converter 16 BIT
Temperature Meter 0 to +100°C with PT1000 probe (0,1°C resolution and ± 0,2°C accuracy)
pH Meter -2,000 to +16,000pH (0,001pH resolution and ± 0,005pH accuracy)
Sampling Interval 200μs
Digital Interface
Connection USB with full instrument control (baud rate 57600 – N – 8 – 1)
Memory EEPROM 64kB – SRAM 32kB
Digital Burette Model 235 with 1000μL calibrated syringe (optional)
Peristaltic Pump Control Start/stop and direct/reverse flow
Cell Connections
Working Electrode Hanging Drop Mercury Electrode (110 x Ø6mm capillary - Ø0,1mm)
Rotating Disk Electrode (100 to 5000rpm)
Solid Electrode Tips
Counter Electrode Platinum wire (Ø0,8 x 5mm) with 2mm ball ending
Reference Electrode Ag/AgCl (3M KCl electrolyte)
Calomel (Sat. KCl electrolyte)
Stirrer Magnetic or overhead propeller (computer controlled)
Power Supply and Dimensions
Voltage Mains 115 or 230V AC ±10% 50/60Hz
Power Consumption 40VA max
Dimensions (L x W x H) 270 x 320 x 350mm
Weight 8kg

Specifications are subject to change without previous notice

For more information visit: https://www.amelchem.com/4330va/