Squidstat Plus ±1A Potentiostat Galvanostat EIS up to 2MHz

±1 A Potentiostat with EIS up to 2 MHz and 1% / 2° Accuracy at 1 mΩ

Squidstat Plus Highlights:

  • Becomes a multichannel system by connecting more Squidstats to the same computer
  • Onboard memory (16 GB) backs up all data collected from experiments for failsafe recovery
  • Voltage/current setpoints, OCP mode, and sampling rates are adjustable in real time
  • Wide selection of prebuilt experiments are available, and the “Drag & Drop” experiment builder can quickly and easily make custom experiments – no coding skills necessary!

Each Squidstat Plus includes one 100 cm channel cable set, one USB cable, and one power supply with over 300 cm in length between the power outlet and potentiostat. The Squidstat User Interface software and unlimited remote technical support are included with purchase.

Description

Channels Per Unit 1
Operating Modes Pot / Gal / ZRA / FRA
Cell Connections 2, 3, 4, or 5 electrode
ADC/DAC Resolution 16 bit resolution
Slew Rate (No Load) 5 MV/s
Maximum Sampling Speed 100,000 samples/s
Floating and/or Earth-grounding Both (Switchable)
Input Impedance > 10 Tera-ohm
Input Bias Current < 1 pA
Data Backup Memory (Onboard) 16 GB
Physical Dimensions 24 cm x 17 cm x 6 cm
Channel Cable Length 100 cm
Computer Interface 1 USB per unit
Power Supply Requirements 100 – 240 VAC, 50 – 60 Hz

Potentiostatic Mode

Compliance Voltage ±12 V per channel
Scan Ranges (Controlled Voltage) 4 ranges (±10 V, ±5 V, ±1 V, ±0.5 V)
Applied Potential Accuracy 0.1% of setpoint, 1 mV min
Applied Potential Resolution 0.003% of range, 15 µV min
Measured Potential Accuracy 0.1% of measurement, 1 mV min
Measured Potential Resolution 0.003% of range, 15 µV min

Galvanostatic Mode

Maximum Current ±1 A
Current Ranges 8 ranges (100 nA to 1 A)
Applied Current Accuracy 0.1% of range, 1 nA min
Applied Current Resolution 0.003% of range, 3 pA min
Measured Current Accuracy 0.1% of range, 1 nA min
Measured Current Resolution 0.003% of range, 3 pA min

Impedance Analyzer

AC Frequency Range for EIS 10 μHz to 2 MHz
AC Frequency Accuracy 0.005% or better
AC Frequency Resolution 0.0004%, 3 µHz max
Potentiostatic Amplitude ≤ 100% of range, 10 V max
Pot. Amplitude Resolution 240 µV, 12 bit resolution
Galvanostatic Amplitude ≤ 100% of range, 1 A max
Gal. Amplitude Resolution 0.002% of range, 2.4 pA max